MALDI-TOF mass spectrometry identification of filamentous fungi in the clinical laboratory

Ranque, S.
Normand, A-C
Cassagne, C
Murat, J.B.
Bourgeois, N.
Dalle, F.
Gari-Toussaint, M.
Fourquet, P.
Hendrickx, Marijke
Piarroux, R.
Citations
Altmetric:
Abstract

This study aimed to validate the effectiveness of a standardised procedure for the MALDI-TOF mass spectrometry (MS)-based identification on a large sample of filamentous fungi routinely identified in university hospitals' laboratories. Non-dermatophyte filamentous fungi prospectively isolated in the routine activity of five teaching hospitals in France were first identified by conventional methods in each laboratory and then by MS in one centre. DNA sequence-based identification resolved discrepancies between both methods. In this study, of the 625 analysed filamentous fungi of 58 species, 501 (80%) and 556 (89%) were correctly identified by conventional methods and MS respectively. Compared with the conventional method, MS dramatically enhanced the performance of the identification of the non-Aspergillus filamentous fungi with a 31

Description
Date
2014-03-01
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
Activity, Clinical, conventional, Dna, filamentous fungi, France, Fungi, hospital, hospitals, identification, Increase, IS, Laboratories, limitation, MALDI-TOF, Mass, Mass Spectrometry, means, method, methods, MS, performance, result, results, routine, routine laboratory, Sample, Species, study, teaching, Universities, university, university hospital, Validate
Citation
Topic(s)
Related project
Embedded videos