Item

Positive interference of merthiolate in the TDx digoxin assay in control samples used in the Belgian External Quality Assessment (EQA) Scheme

Devleeschouwer, N.
Libeer, J.C.
Schiettecatte, J.
Maes, Virginie
Demolder, J.
Debusschere, P.M.
Gorus, F.
Citations
Altmetric:
Abstract

We investigated the possible origin of the spuriously high results observed with the Abbott TDx Immunoassay in the 1991 Belgian external quality assessment scheme for digoxin. The present work ascribes this discrepancy to a matrix effect induced by the addition of merthiolate as preservative to the control samples. It consequently stresses the importance of avoiding the use of this compound for preparing such samples

Description
Date
1993-12-12
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
a, article, AS, assessment, Belgian, Belgium, Biology, blood, Brussels, Clinical, Comparative Study, Control, Digoxin, effect, epidemiology, EQA, external, External quality assessment, External quality assessment scheme, False Positive Reactions, Fluorescence Polarization Immunoassay, Humans, Hydrocortisone, Hygiene, im, Immunoassay, Interference, IS, IT, journal, matrix, observed, ORIGIN, present, Print, Progesterone, Quality, Quality Control, Radioimmunoassay, Reagent Kits,Diagnostic, result, results, Sample, Samples, SB - IM, stress, study, Thimerosal, use, work