Loading...
Thumbnail Image
Item

Fourier transform infrared immunosensors for model hapten molecules.

Gosselin, E
Gorez, M
Voué, M
Denis, Olivier J.
Conti, J
Popovic, N
Van Cauwenberg, A
Noel, E
De Coninck, J
Citations
Altmetric:
Abstract

We report experimental results concerning the detection of 2,4-dinitrophenol, under its free form or coupled to human serum albumin using Fourier transform infrared spectroscopy-based sensors. Competitive immunoreactions were carried out using several anti-dinitrophenol monoclonal antibodies. Comparison with enzyme-linked immunosorbent assays in competition is given for standard operating conditions. FTIR detection limits are comparable to those obtained by ELISA. The limits of detection are about 5-15 ng/mL for the coupled DNP. Using the LO-DNP61 antibody, a detection limit of congruent with 5 ng/mL was also estimated for the free DNP molecules but is much higher for the other antibodies.

Description
Date
2009-04-15
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
Antibodies, Monoclonal, Biosensing Techniques, Equipment Design, Equipment Failure Analysis, Haptens, Immunoassay, Reproducibility of Results, Sensitivity and Specificity, Spectroscopy, Fourier Transform Infrared