Item

Optimization and validation of a micellar electrokinetic chromatographic method for the analysis of several angiotensin-II-receptor antagonists

Hillaert, S.
De Beer, T.R.
De Beer, J.
Van den Bossche, W.
Citations
Altmetric:
Abstract

We have optimized a micellar electrokinetic capillary chromatographic method for the separation of six angiotensin-II-receptor antagonists (ARA-IIs): candesartan, eprosartan mesylate, irbesartan, losartan potassium, telmisartan, and valsartan. A face-centred central composite design was applied to study the effect of the pH, the molarity of the running buffer, and the concentration of the micelle-forming agent on the separation properties. A combination of the studied parameters permitted the separation of the six ARA-IIs, which was best carried out using a 55-mM sodium phosphate buffer solution (pH 6.5) containing 15 mM of sodium dodecyl sulfate. The same system can also be applied for the quantitative determination of these compounds, but only for the more stable ARA-IIs (candesartan, eprosartan mesylate, losartan potassium, and valsartan). Some system parameters (linearity, precision, and accuracy) were validated

Description
Date
2003-01-10
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
Agent, Agents, analysis, Angiotensin II, Angiotensin Receptor Antagonists, Antihypertensive Agents, article, Belgium, Buffers, chemistry, Chromatography,Micellar Electrokinetic Capillary, Combination, Design, DRUG, effect, Faculty, Hydrogen-Ion Concentration, im, IS, journal, Laboratories, metabolism, method, methods, parameters, ph, pharmacology, Potassium, Print, receptors, Receptors,Angiotensin, Reference Standards, Reproducibility of Results, SB - IM, Science, separation, sodium, study, System, Universities, university, VALIDATION, Validation study
Citation
Topic(s)
Related project
Embedded videos