Item

Interference of dextran in biuret-type assays of serum proteins

Delanghe, J.R.
Hamers, N.
Taes, Y.E.
Libeer, J.C.
Citations
Altmetric:
Abstract

Dextran interference in biuret-type assays of total serum proteins was investigated in a Belgian National External Quality Assurance Survey with 256 participants. In vitro supplementation of therapeutic (10% Gentran 70) dextran concentrations showed a broadly varying (from 0 to 20%) negative interference. The analytical interference was found to depend on both the sodium hydroxide and tartrate concentrations in the reagent formulation. The dry chemistry biuret method was not affected by the dextran interference. In a number of cases, the effects observed may be of clinical importance. Both clinicians and laboratory staff should be aware of the persistence of this analytical problem

Description
Date
2005-07-01
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
a, analysis, article, Belgian, Belgium, Biuret Reaction, blood, Blood Proteins, Case, chemistry, Clinical, Clinical Chemistry Tests, Dextrans, effect, effects, external, hospital, Humans, im, Interference, IS, journal, Laboratories, method, methods, national, observed, PARTICIPANTS, persistence, plasma, Plasma Substitutes, Print, protein, Proteins, Quality, QUALITY ASSURANCE, Reagent Kits,Diagnostic, regression analysis, SB - IM, serum, sodium, Supplementation, survey, Universities, university, university hospital
Citation
Topic(s)
Related project
Embedded videos