Item

Factors influencing between-laboratory variability of C-reactive protein results as evidenced by the Belgian External Quality Assessment(EQA) Scheme

Devleeschouwer, N.
Libeer, J.C.
Chapelle, J.P.
Struway, C.L.
Gyssels, C.
L'Hoir, A.
Albert, A.
Citations
Altmetric:
Abstract

Based on results from the Belgian External Quality Assessment (EQA) Scheme, we studied the main factors affecting the between-laboratory variation of C-reactive protein determination. Participants using homogeneous systems with several calibration points generally achieved better performance. Working temperatures influenced the results to a lesser extent. The present study stresses the importance for EQA organizers to collect more detailed information about CRP analytical methods used by the participants. It also suggests that manufacturers should be more involved in the management of quality, in particular by striving for standardization of the material (kit and calibrator) they produce for CRP assay

Description
Date
1993-12-10
Journal Title
Journal ISSN
Volume Title
Publisher
Chapter title
Publication type
Peer reviewed scientific article
Research Projects
Organizational Units
Journal Issue
Keywords
analysis, article, AS, assessment, Belgian, Belgium, biologie clinique, Brussels, C-Reactive Protein, Calibration, Comparative Study, de, EQA, ET, external, External quality assessment, factors, Fluorescence Polarization Immunoassay, Humans, im, Immunodiffusion, INFORMATION, IS, IT, journal, Laboratories, management, method, methods, Nephelometry and Turbidimetry, PARTICIPANTS, performance, present, Print, protein, Quality, Quality Control, Reproducibility of Results, result, results, SB - IM, Service, STANDARDIZATION, standards, stress, study, System, Systems, Temperature, VARIABILITY, variation
Citation
Topic(s)
Related project
Embedded videos